Roswell – Autoclavable

Roswell - Autoclavable

Roswell – Autoclavable Version

Description
Dimensions/tolerance(in) 1.89 x 1.10 x 0.53 (+/- 0.02)
Dimensions/tolerance (mm) 48 x 28 x 13.5 (+/- 0.5)
IP rating IP69K
Attachment Welding, cable tie, rivet hole ø 0.19 in (ø 5 mm)
Application temperature -40°F to + 482°F (-40°C to + 250°C)
Operating temperature -40°F to + 185°F (-40°C to + 85°C)
Read range on metal, 4W EIRP (2W ERP) Up to 16.4 ft (5 m)
Weight / tolerance (w/o bracket) 0.92 oz / 26 g (+/- 10%)
Compliance FDA Class I, RoHS, CE
Dimensions w/o bracket/ tolerance (mm) ø 28 x 11 (+/- 0.5)
Dimensions w/o bracket/ tolerance (in) ø 1.10 x 0.43 (+/- 0.02)
Weight / tolerance 1.55 oz / 44 g (+/- 10%)
Material Stainless steel

Part Number

  • Autoclavable Roswell w/ Bracket US: X1114-US141-H3
  • Autoclavable Roswell w/ Bracket EU: X1114-EU141-H3
  • Autoclavable Roswell w/o Bracket US: X1114-US140-H3
  • Autoclavable Roswell w/o Bracket EU: X1114-EU140-H3

Roswell, the first tag of its kind, uses only its stainless steel interior as an antenna. The patent-pending tag is constructed of two cylindrical metal cases with a thin high temperature substrate between them. While existing products in the market may use a tag packaged inside a metal case, Roswell is completely different. Because there is no RFID tag packaged inside Roswell, the antenna-free design ensures reliability and durability by eliminating the risk of tag failure from high temperature and high pressure autoclave and corrosive liquids. Roswell is designed for quick an convenient attachment and can be welded directly to metal trays or easily attached via a stainless steel bracket to a medical equipment.

Functional Specifications
RF air protocol EPC Class 1 Gen 2; ISO18000-6C
Operating frequency UHF 902-928 MHz (US); 866-868 MHz (EU)
IC type Alien Higgs-3
Memory configuration 96-bit EPC ; 512-bit user memory; 64-bit TID
Functionality Read / write
Memory – expected read / write cycles 100,000 cycles at 77°F (25°C)
Data retention Up to 50 years
Read rate 400 tags per second for 96-EPC bit number
Warrenty (limited) 1 year

 

Performance Characteristics
Read range on metal (2W ERP) Up to 16.4 ft (5 m)
Polarization Linear

 

Physical Specifications
Material Stainless steel
Attachment Welding, cable tie, rivet hole ø 0.19 in (ø 5 mm)
Color Stainless steel

 

Environmental and Industry Compliance
RoHS EU Directive 2002/95/EC
ATEX ATEX Compliant
Healthcare sterilization compliance ISO 17665 – Moist Steam, ISO 11135 – Ethylene Oxide
FDA class Class I

 

Cleaning and Disinfection Process (ANSI/AAMI ST79)
Chemical sterilization for decontamination Liquid chemical sterilants include glutaraldehyde, hydrogen peroxide, hydrogen peroxide-peracetic acid combinations, sodium hypochlorite, and peracetic acid Pass
Thermal (hot water) disinfection

• Washer-sanitizers, washer-decontaminators, utensil washers, and cart washers reduce microbial contamination by means of cleaning agents, hot water, rinsing, and drying.

• Pasteurization equipment provides cleaning and high-level disinfection at water temperatures of 65°C to 77°C (150°F to 170°F) for a contact time of at least 30 minutes.

• Washer-disinfectors provide a cycle of cleaning, rinsing, disinfection, and drying at temperatures that are usually higher than those of washer-sanitizers

Pass
Thermal sterilization for decontamination Saturated steam can be used to decontaminate devices capable of withstanding high temperatures (121°C to 135°C [250°F to 275°F]) and pressures (16 to 35 psig). Washer-sterilizers provide cleaning and rinsing followed by exposure to saturated steam at temperatures of 121°C to 135°C (250°F to 275°F). Pass
Ultrasonic cleaner 3KHz for 30 min Pass
Dry heat sterilization 160°C for 2h, or 170°C for 1h, or 180°C for 30min Pass
Disinfection by light C wave ultraviolet and wavelength range is from 200nrn to 275nrn Pass
Detergent enzyme cleaner Pass
Autoclave sterilization 121°C and 115Kpa for 1000 cycles Pass

 

Operational and Environmental Specifications
Operational temperature
Cold -40°F (-40°C)
Dry heat (Long term, days/weeks/years) +185°F (+85°C)
Application temperature
Cold -40°F (-40°C)
Dry heat (Short term, minutes/hours) +482°F (+250°C)
Temperature cycling testing 6 hours at +250°C, 18 hours cool-down, 30 days test cycle
Humidity
Operational humidity 5%-95% non-condensing
Storage humidity 5%-95% non-condensing
Compression strength 1145 psi (7900 kPa)
Drop 3 ft (1 m) to concrete/granite up to 100 cycle
Vibration MIL-STD-810F
IP classification IP69K

 

Product Dimensions and Weight
Dimensions / tolerance(mm) 48 x 28 x 13.5 (+/- 0.5)
Dimensions / tolerance (in) 1.89 x 1.10 x 0.53 (+/- 0.02)
Weight / tolerance 1.55 oz / 44 g (+/- 10%)
Dimensions w/o bracket / tolerance (mm) ø 28 x 11 (+/- 0.5)
Dimensions w/o bracket / tolerance (in) ø 1.10 x 0.43 (+/- 0.02)
Weight / tolerance (w/o bracket) 0.92 oz / 26 g (+/- 10%)

Specialities

  • Withstands repeated autoclave cycles
  • Resistant to chemical cleaning process
  • Designed to meet FDA requirements
  • 180 degree read orientation

Applications

  • Autoclavable Surgical Trays
  • Medical devices

Add Comment