Roswell – Autoclavable Version
Description | |
---|---|
Dimensions/tolerance(in) | 1.89 x 1.10 x 0.53 (+/- 0.02) |
Dimensions/tolerance (mm) | 48 x 28 x 13.5 (+/- 0.5) |
IP rating | IP69K |
Attachment | Welding, cable tie, rivet hole ø 0.19 in (ø 5 mm) |
Application temperature | -40°F to + 482°F (-40°C to + 250°C) |
Operating temperature | -40°F to + 185°F (-40°C to + 85°C) |
Read range on metal, 4W EIRP (2W ERP) | Up to 16.4 ft (5 m) |
Weight / tolerance (w/o bracket) | 0.92 oz / 26 g (+/- 10%) |
Compliance | FDA Class I, RoHS, CE |
Dimensions w/o bracket/ tolerance (mm) | ø 28 x 11 (+/- 0.5) |
Dimensions w/o bracket/ tolerance (in) | ø 1.10 x 0.43 (+/- 0.02) |
Weight / tolerance | 1.55 oz / 44 g (+/- 10%) |
Material | Stainless steel |
Part Number
- Autoclavable Roswell w/ Bracket US: X1114-US141-H3
- Autoclavable Roswell w/ Bracket EU: X1114-EU141-H3
- Autoclavable Roswell w/o Bracket US: X1114-US140-H3
- Autoclavable Roswell w/o Bracket EU: X1114-EU140-H3
Roswell, the first tag of its kind, uses only its stainless steel interior as an antenna. The patent-pending tag is constructed of two cylindrical metal cases with a thin high temperature substrate between them. While existing products in the market may use a tag packaged inside a metal case, Roswell is completely different. Because there is no RFID tag packaged inside Roswell, the antenna-free design ensures reliability and durability by eliminating the risk of tag failure from high temperature and high pressure autoclave and corrosive liquids. Roswell is designed for quick an convenient attachment and can be welded directly to metal trays or easily attached via a stainless steel bracket to a medical equipment.
Functional Specifications | |
---|---|
RF air protocol | EPC Class 1 Gen 2; ISO18000-6C |
Operating frequency | UHF 902-928 MHz (US); 866-868 MHz (EU) |
IC type | Alien Higgs-3 |
Memory configuration | 96-bit EPC ; 512-bit user memory; 64-bit TID |
Functionality | Read / write |
Memory – expected read / write cycles | 100,000 cycles at 77°F (25°C) |
Data retention | Up to 50 years |
Read rate | 400 tags per second for 96-EPC bit number |
Warrenty (limited) | 1 year |
Performance Characteristics | |
---|---|
Read range on metal (2W ERP) | Up to 16.4 ft (5 m) |
Polarization | Linear |
Physical Specifications | |
---|---|
Material | Stainless steel |
Attachment | Welding, cable tie, rivet hole ø 0.19 in (ø 5 mm) |
Color | Stainless steel |
Environmental and Industry Compliance | |
---|---|
RoHS | EU Directive 2002/95/EC |
ATEX | ATEX Compliant |
Healthcare sterilization compliance | ISO 17665 – Moist Steam, ISO 11135 – Ethylene Oxide |
FDA class | Class I |
Cleaning and Disinfection Process (ANSI/AAMI ST79) | |
---|---|
Chemical sterilization for decontamination Liquid chemical sterilants include glutaraldehyde, hydrogen peroxide, hydrogen peroxide-peracetic acid combinations, sodium hypochlorite, and peracetic acid | Pass |
Thermal (hot water) disinfection
• Washer-sanitizers, washer-decontaminators, utensil washers, and cart washers reduce microbial contamination by means of cleaning agents, hot water, rinsing, and drying. • Pasteurization equipment provides cleaning and high-level disinfection at water temperatures of 65°C to 77°C (150°F to 170°F) for a contact time of at least 30 minutes. • Washer-disinfectors provide a cycle of cleaning, rinsing, disinfection, and drying at temperatures that are usually higher than those of washer-sanitizers |
Pass |
Thermal sterilization for decontamination Saturated steam can be used to decontaminate devices capable of withstanding high temperatures (121°C to 135°C [250°F to 275°F]) and pressures (16 to 35 psig). Washer-sterilizers provide cleaning and rinsing followed by exposure to saturated steam at temperatures of 121°C to 135°C (250°F to 275°F). | Pass |
Ultrasonic cleaner 3KHz for 30 min | Pass |
Dry heat sterilization 160°C for 2h, or 170°C for 1h, or 180°C for 30min | Pass |
Disinfection by light C wave ultraviolet and wavelength range is from 200nrn to 275nrn | Pass |
Detergent enzyme cleaner | Pass |
Autoclave sterilization 121°C and 115Kpa for 1000 cycles | Pass |
Operational and Environmental Specifications | |
---|---|
Operational temperature | |
Cold | -40°F (-40°C) |
Dry heat (Long term, days/weeks/years) | +185°F (+85°C) |
Application temperature | |
Cold | -40°F (-40°C) |
Dry heat (Short term, minutes/hours) | +482°F (+250°C) |
Temperature cycling testing | 6 hours at +250°C, 18 hours cool-down, 30 days test cycle |
Humidity | |
Operational humidity | 5%-95% non-condensing |
Storage humidity | 5%-95% non-condensing |
Compression strength | 1145 psi (7900 kPa) |
Drop | 3 ft (1 m) to concrete/granite up to 100 cycle |
Vibration | MIL-STD-810F |
IP classification | IP69K |
Product Dimensions and Weight | |
---|---|
Dimensions / tolerance(mm) | 48 x 28 x 13.5 (+/- 0.5) |
Dimensions / tolerance (in) | 1.89 x 1.10 x 0.53 (+/- 0.02) |
Weight / tolerance | 1.55 oz / 44 g (+/- 10%) |
Dimensions w/o bracket / tolerance (mm) | ø 28 x 11 (+/- 0.5) |
Dimensions w/o bracket / tolerance (in) | ø 1.10 x 0.43 (+/- 0.02) |
Weight / tolerance (w/o bracket) | 0.92 oz / 26 g (+/- 10%) |
Specialities
- Withstands repeated autoclave cycles
- Resistant to chemical cleaning process
- Designed to meet FDA requirements
- 180 degree read orientation
Applications
- Autoclavable Surgical Trays
- Medical devices